5.1.2 Changing the wax

5.Operation

5.1.2 Changing the wax

1

Draining the wax

Connect the paraffin drain hose (2) as shown in Fig. 13 and insert the loose end of the hose into a remote collecting container.

→ +

2

Fig. 13

Make sure the hose end is securely kept in the remote container until both the drain has been completed and the hose has been purged with pressurized air.

From the FAVORITES screen, press SMART SCREEN (see previous chapter).

Select the wax bath to be drained and the remote container (‘Remote’).

Then press STATION DRAIN.

Several queries will be displayed each of which has to be confirmed individually.