Leica ASP300 instruction manual Operation, Program Options

Models: ASP300

1 70
Download 70 pages 22.12 Kb
Page 35
Image 35
Program Options

5. Operation

Program Options

Selection of the options on how programs are carried out.

STATION ALLOCATION:

By age or Sequential

 

By age -

RMS is activated, reagents will automatically be

 

 

used in order of increasing cleanliness.

 

Sequential -

reagents will be used in sequential order of stations.

WAX BATH ORDER:

Auto or 1.; 2. ; 3.

Wax bath order can be activated only when STATION ALLOCATION is set to Sequential.

• PROMPT FOR NUM BLOCKS :

Enabled or Disabled

Enabled- When starting a program, the actual number of blocks must be entered.

This is recommended with the RMS being enabled.

Disabled - Number of blocks can be entered optionally (not mandatory).

Date / Time

Be sure to verify that the date and time entered do actually correspond to the local time/date.

Press the buttons to access the entry windows where time/date can be corrected, if necessary.

Instrument

Press WAX BATH SET TEMP (°C) to access the wax bath temperature entry window. Select the highest temperature compatible with the type of paraffin used as paraffin temperature will drop slightly once the paraf- fin is in the retort.