Leica ASP300 instruction manual Operation, Entering / modifying reagent thresholds

Models: ASP300

1 70
Download 70 pages 22.12 Kb
Page 38
Image 38
Entering / modifying reagent thresholds

5. Operation

 

 

Allocation of reagents to the correct reagent group is the basis for

 

 

compatibility monitoring. Allocation to the wrong reagent group

 

 

can lead to reagent cross contamination.

Changeable parameters

 

 

Entering / modifying reagent thresholds

 

• Highlight the reagent to be modified.

 

• Press the headline key symbol of the parameter to be changed.

 

• Enter a new threshold or - if no warning is desired - press CLEAR to

 

remove the threshold altogether.

 

• Press OK to confirm.

 

 

 

 

 

The same threshold value applies for all reagent stations con-

 

 

taining the same reagent.

 

 

 

Changing reagent names or reagent groups

Highlight the reagent.

Press the corresponding headline key symbol.

In the entry window (or via the keyboard), enter the new reagent group allocation.

Press OK to confirm and save.

Reedit all station and programs allocated to the reagent changed.

38