5.4Reagent status

5.Operation

5.4Reagent status

In the reagent status screen (Fig. 32) you can display and update the actual reagent status of each station and you can set the stations as full or empty.

Warning thresholds that trigger warning messages appear on a red background.

To change a reagent status, highlight the corresponding line and press the corresponding key symbol .

Fig. 32

Reagent warnings

If the RMS is enabled, the instrument will give you warning messages when reagents have been used for too long.

Warning messages for normal reagents are given based on up to 3 criteria; warning mes- sages for paraffin wax on up to 6 criteria.

The warning messages are displayed at the end of a cleaning cycle. From there, you can go directly to the REAGENT STATUS screen.

In addition, warning messages (on a yellow background) are displayed in the program start screens.

Criteria for warning messages

For normal reagents, the following warning cri- teria apply:

number of blocks processed

number of programs carried out

number of day lapsed

since the reagent has last been changed.

For paraffin wax, three additional criteria can be applied:

number of blocks processed

number of programs carried out

number of days lapsed

since the wax has last been cleaned.

52

Instruction Manual V 1.3 – 12/2003

Page 52
Image 52
Leica ASP300 instruction manual 5.4Reagent status →, Operation