Leica ASP300 Cleaning and maintenance, Draining the condensate bottle, Filter label

Models: ASP300

1 70
Download 70 pages 22.12 Kb
Page 64
Image 64
Draining the condensate bottle

7.Cleaning and maintenance

Draining the condensate bottle

• Remove and empty the condensate bottle and reinsert (Fig. 39).

Fig.39

Exchanging the activated carbon filter

• Exchange the filter as shown in Fig. 40.

• Replace the filter at least every four months

– make sure to insert correctly – (see arrow on label)!

• Perform a Smart Clean prior to changing the filter.

Fig. 40

Filter label

64

Instruction Manual V 1.3 – 12/2003

Page 64
Image 64
Leica ASP300 Cleaning and maintenance, Draining the condensate bottle, Exchanging the activated carbon filter