5.Operation
Access the RUN OPTIONS function to display the options selected for the current program.
• | LEVEL CHECK enabled: |
↓ | The lower level sensor in the retort is acti- |
vated. | |
• | RECIRCULATION enabled: |
| During each processing step, the reagent |
| is pumped periodically from the bottle to the |
| retort and back for continuous and efficient |
| mixing of fluid. The first pump step is carried |
| out after 16 minutes, all subsequent pump |
| steps in |
• | PRETEST enabled: |
| After starting the program a test run is per- |
| formed: the first reagent is pumped into the |
| retort and back into the bottle. |
• | STIRRER enabled: |
| Magnetic stirrer is on. |
• | PRINTER enabled: |
| At the end of the program run, a full run log |
| will be printed out. |
Fig. 30
Important!
The options selected will be applied to ALL programs!
All run options can be enabled / disabled during the program run if necessary.
The printer option will be displayed only if a printer has been configured for the instrument.
50 | Instruction Manual V 1.3 – 12/2003 |