LEVEL CHECK enabled:

5.Operation

Access the RUN OPTIONS function to display the options selected for the current program.

LEVEL CHECK enabled:

The lower level sensor in the retort is acti-

vated.

RECIRCULATION enabled:

 

During each processing step, the reagent

 

is pumped periodically from the bottle to the

 

retort and back for continuous and efficient

 

mixing of fluid. The first pump step is carried

 

out after 16 minutes, all subsequent pump

 

steps in 12-minute intervals.

PRETEST enabled:

 

After starting the program a test run is per-

 

formed: the first reagent is pumped into the

 

retort and back into the bottle.

STIRRER enabled:

 

Magnetic stirrer is on.

PRINTER enabled:

 

At the end of the program run, a full run log

 

will be printed out.

Fig. 30

Important!

The options selected will be applied to ALL programs!

All run options can be enabled / disabled during the program run if necessary.

The printer option will be displayed only if a printer has been configured for the instrument.

50

Instruction Manual V 1.3 – 12/2003

Page 50
Image 50
Leica ASP300 instruction manual Operation, LEVEL CHECK enabled