Omega Speaker Systems FPU5-MT-220 2cont’d, Flow Rate Mode, Display, Action Taken, Shows, 6-10

Models: FPU5-MT-220 FPU5-MT-110

1 83
Download 83 pages 7.89 Kb
Page 47
Image 47
Table 6-2 (cont’d)

 

Table 6-2 (cont’d)

 

FLOW RATE MODE

 

KEY

DISPLAY

 

ACTION TAKEN

 

 

 

SHOWS

 

AND PROCEDURE

 

 

 

 

 

 

 

 

 

xx:xx:xx

 

Displays cumulative TIME since start

 

 

TIME

 

 

 

 

 

 

Press the TIME key. The display flashes the time that the pump has

 

 

 

 

 

 

 

 

been running (xx:xx:xx).

 

 

 

 

 

 

 

 

 

xxx mL/min

 

Displays current FLOW RATE setting

 

FLOW

 

 

 

RATE

 

or

 

 

 

 

 

1.

Press the FLOW RATE key. The display flashes the flow rate

 

 

 

 

 

 

xxx L/min

 

 

 

 

setting you selected just before starting the pump.

 

 

 

 

 

2.Use the UP and DOWN ARROW keys to enter the FLOW RATE setting you desire. As you use the UP and DOWN ARROW keys, the display changes accordingly. Refer to Figure 6-4.

 

 

 

 

 

 

xxx mL/min

Displays current TOTAL VOLUME

 

 

 

 

TOTAL

 

 

 

 

VOLUME

 

or

 

 

 

 

 

 

 

xxx Liters

Press the TOTAL VOLUME key. The display flashes the calculated

 

 

 

 

 

 

 

volume of fluid dispensed since the pump started running (xxx mL or

 

 

 

 

 

 

 

xxx Liters).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TUBE ID=xxx

Displays current TUBE ID

 

 

TUBE ID

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the TUBE ID key. If the red LED next to the CALIBRATE key

 

 

 

 

 

 

 

 

 

 

 

 

 

 

is not lit, the display flashes the current TUBE ID setting. If the red

 

 

 

 

 

 

 

LED next to the CALIBRATE key is lit, the display flashes the

 

 

 

 

 

 

 

message “CALIBRATED TUBE!”.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CALIBRATED!

Displays calibration state

 

 

 

CALIBRATE

 

 

 

 

 

or

 

 

 

 

 

 

 

 

 

 

 

 

 

 

not calibrated

1. Press the CALIBRATE key. If the red LED next to the

 

 

 

 

 

 

 

CALIBRATE key is not lit, the display shows “not calibrated”..

 

 

 

 

 

 

 

If the red LED next to the CALIBRATE key is lit, the display

 

 

 

 

 

 

 

shows “CALIBRATED!”.

 

 

 

 

 

 

 

To calibrate a tube:

2.Press the START/STOP key to stop the pump. The display returns to the STOPPED Display Mode. For detailed instruc- tions on “Calibrating a Tube”, refer to Section 5.5.

 and 

xxx mL/min

NOTE: if you press the UP and DOWN ARROW keys, the display

 

or

shows the current FLOW RATE setting. This setting

 

xxx L/min

changes accordingly as you use the UP and DOWN

 

 

ARROW keys but does not wrap around at maximum and

 

 

minimum values while the pump is running.

 

 

 

Language

 

NOTE: if you press the Language button, the display changes to

Button

 

the next language in the sequence: English, French,

 

 

Spanish, German, and Italian.

6-10

Page 47
Image 47
Omega Speaker Systems FPU5-MT-220, FPU5-MT-110 2cont’d, Flow Rate Mode, Display, Action Taken, Shows, And Procedure, 6-10