Image 54
6.4.1 Pumping in the TOTAL VOLUME Mode

Operating the Pump Motor in the RUNNING Display Mode

6

 

 

6.4.1 Pumping in the TOTAL VOLUME Mode

To pump fluid in the TOTAL VOLUME Mode, do the following:

1.Set up the pump motor for pumping, according to Chapter 3.

2.Toggle the ON/OFF switch to turn on the unit’s power. The display is in the STOPPED Display Mode. Refer to Figure 6-7.

3.Using the procedure outlined in Sections 6.1.1. and 6.2.1, adjust either the RPM or FLOW RATE setting.

4.When the display flashes the RPM or FLOW RATE setting you desire, press the TOTAL VOLUME key.

TOTAL

VOLUME

Display shows

xxx mL or xxx Liters

ACTION

Displays current

TOTAL VOLUME

setting

The display flashes the current TOTAL VOLUME setting (xxx mL or xxx Liters).

5.Use the UP and DOWN ARROW keys to enter the TOTAL VOLUME setting you desire. Refer to Figure 6-8 for the TOTAL VOLUME range.





Display shows

Display shows

xxx mL or xxx Liters

xxx mL or xxx Liters

ACTION

ACTION

TOTAL VOLUME setting scrolls up

TOTAL VOLUME setting scrolls down

Range

Range





1 mL

10,000 L

 

2 mL

 

 

 

9,999 L

 

.

 

 

 

.

 

999 mL

 

 

 

1,000 mL (1L)

1,000 mL (1 L)

 

 

 

999 mL

 

 

 

.

 

 

 

.

 

9,999 L

 

 

 

 

 

 

2 mL

 

10,000 L

 

 

 

 

1 mL

 

 

 

 

Figure 6-8. TOTAL VOLUME Range

6-17

Page 54
Image 54
Omega Speaker Systems FPU5-MT-110, FPU5-MT-220 manual Pumping in the TOTAL VOLUME Mode