Omega Speaker Systems FPU5-MT-110 2.Average Flow Rates, Tubing Size, Due to, Between Pumps, Motor

Models: FPU5-MT-220 FPU5-MT-110

1 83
Download 83 pages 7.89 Kb
Page 74
Image 74
Table 9-2. Average Flow Rates

9Specifications

Flow rate tests were done with 20°C water at 0 PSI back pressure.

 

 

Table 9-2. Average Flow Rates

 

 

 

 

 

 

Tubing Wall

Tubing Size

mL per

Minimum Flow

Maximum Flow

Thickness

Inner Diameter

Revolution

Rate at 600 RPM

Rate at 600 RPM

 

 

 

(mL/Minute)

(mL/Minute)

116"

132"

0.05

1

30

116"

116"

0.22

3

132

116"

18"

0.9

9

540

116"

316"

1.9

19

1140

116"

14"

3.0

30

1800

116"

516"

3.8

38

2280

1.5 mm

1.0 mm

0.08*

1

48

1.5 mm

2.0 mm

0.35*

4

210

1.5 mm

3.0 mm

0.8*

8

482

1.5 mm

4.0 mm

1.43*

15

857

1.5 mm

5.0 mm

2.1*

21

1257

1.5 mm

6.0 mm

2.7*

27

1607

1.5 mm

7.0 mm

3.6*

37

2187

1.5 mm

8.0 mm

3.9*

39

2316

* Metric mL per revolution numbers are rounded off

To determine the chemical compatibility of a particular tubing material, it is recommended that you test the tubing under actual conditions

Tubing materials that can be used include Vinyl, Viton, Tygon, Silicone, Santoprene, and Norprene. Up to 68 durometer tubing can be used.

Table 9-3. Variances in Normalized Flow Rate (mL per Revolution)

Tubing Size

Due to

Between Pumps

Due to Different

Inner Diameter

Motor

(due to Mechanical

Tubing Materials

(in.)

Speed

Tolerances)

 

 

 

 

 

132

±10%

±25%

±20%

116

±10%

±15%

±15%

18

±10%

±10%

±5%

316

±5%

±10%

±5%

14

±5%

±15%

±5%

516

±15%

±20%

±10%

9-2