Table 6-4 TOTAL VOLUME MODE
|
| KEY |
| DISPLAY | ACTION TAKEN | ||
|
|
|
|
| SHOWS | AND PROCEDURE | |
|
|
|
|
|
|
|
|
| ON/OFF Switch |
| xxx mL | Pump motor’s power | |||
|
|
|
| ||||
|
|
|
|
| or | NOTE: if you turn off the pump motor’s power, the pump stops | |
|
|
|
|
| xxx Liters | ||
|
|
|
|
| running. | ||
|
|
|
|
|
| ||
|
|
|
|
|
| To continue pumping, turn on the unit’s power. The display | |
|
|
|
|
|
| returns to the STOPPED Display Mode and shows the volume of | |
|
|
|
|
|
| fluid dispensed during pumping. Then press the START/STOP | |
|
|
|
|
|
| key. | |
|
|
|
|
|
|
|
|
|
|
|
|
| xxx mL | Pump(s) stops; press again pump(s) starts | |
|
|
| START |
| |||
|
|
|
|
|
| ||
|
|
| STOP |
| or |
|
|
|
|
|
|
| To resume pumping for the remainder of the TOTAL VOLUME | ||
|
|
|
|
| xxx Liters | setting, press the START/STOP key. The display enters the | |
|
|
|
|
|
| RUNNING Display Mode, and the fluid begins to dispense. | |
|
|
|
|
|
|
|
|
|
|
|
|
| Must STOP first! | LEDs indicate new pump direction | |
|
| INLET |
|
| |||
|
|
|
|
|
| ||
|
| SELECT |
|
|
|
|
|
|
|
|
|
|
| To select the other inlet and reverse the direction of pumping: | |
|
|
|
|
|
|
1.Press the START/STOP key. The display returns to the STOPPED Display Mode.
2.Press the INLET SELECT key to reverse the direction of pumping.
3.Press the START/STOP key again to resume pumping.
4.For detailed instructions on “Defining the Direction of Pumping, refer to Section 5.2.
|
| STOP to Prime... | Pumps fluid at current rpm or flow rate |
| PRIME | setting as long as the key is pressed | |
|
| ||
|
|
|
|
To prime a tube:
1.Press the START/STOP key to stop the pump. The display returns to the STOPPED Display Mode.
2.Press and hold down the PRIME key to prime the tube.
3.Press the START/STOP key again to resume pumping.
4.For detailed instructions on “Priming a Tube”, refer to Section 5.4.
xxx °C | Display alternates when pressed, °C then °F. |
TEMP
NOTE: The TEMP key functions the same whether the pump is stopped or running. Refer to Section 5.6.
| Displays actual RPM |
RPM | xxx rpm |
| Press the RPM key. The display flashes the actual pump motor |
| speed (xxx rpm). |
| Displays cumulative TIME since start |
TIME | xx:xx:xx |
Press the TIME key. The display flashes the time that the pump has been running (xx:xx:xx:).