Omega Vehicle Security FPU5-MT-110, FPU5-MT-220 manual Specifications, 1.Calibration Accuracy

Models: FPU5-MT-110 FPU5-MT-220

1 83
Download 83 pages 6.26 Kb
Page 73
Image 73
9 Specifications

9 Specifications

Maximum Fluid Back Pressure:

20 PSI

Max Suction Lift:

20 feet of H2O (6.1 meters of H2O)

Temperature Range (Fluid):

-50°F to 300°F (-46°C to -149°C)

Flow Rate:

1 to 2280 mL/min

Motor:

Reversible, 75 W (110 HP), 90 Vdc (10 to

 

600 rpm)

Display:

16 character dot matrix LCD, 9 mm (38") high

 

characters

Power:

90 to 130 Vac, 1.5 A; 190 to 260 Vac, 0.8 A

 

(Unit may be used with extension cord)

Line cord:

1.8 m (6 ft) with US standard plug on 115 Vac

 

units 1.8 m (6 ft) with European plug on

 

230 Vac units

Housing:

Painted sheet metal with rubber feet; units are

 

stackable

Dimensions (H x W x D):

152 x 254 x 254 mm (6" x 10" x 10")

Weight:

8.2 kg (18 lb)

Fuse:

2 amp, 110 Vac or 1 amp, 220 Vac

 

 

Table 9-1. Calibration Accuracy

 

 

 

 

 

 

 

 

Tubing Wall

Tubing Size

Calibration

Tubing Wall

Tubing Size

Calibration

Thickness

Inner Diameter

Accuracy

Thickness

Inner Diameter

Accuracy

116"

132"

±4%

1.5 mm

1.0 mm

±4%

 

 

 

 

 

 

116"

116"

±2%

1.5 mm

2.0 mm

±2%

 

 

 

 

 

 

116"

18"

±2%

1.5 mm

3.0 mm

±2%

 

 

 

 

 

 

116"

316"

±2%

1.5 mm

4.0 mm

±2%

 

 

 

 

 

 

116"

14"

±2%

1.5 mm

5.0 mm

±2%

 

 

 

 

 

 

116"

516"

±4%

1.5 mm

6.0 mm

±2%

 

 

 

 

 

 

 

 

 

1.5 mm

7.0 mm

±4%

 

 

 

 

 

 

 

 

 

1.5 mm

8.0 mm

±4%

 

 

 

 

 

 

NOTE

Calibration accuracy studies were done by calibrating the pump motor and comparing actual volume pumped to requested total volume. Various tube materials were used along with 20°C water, 0 PSI back pressure.

Accuracies may vary with fluid viscosity, temperature and back pressure.

9-1

Page 73
Image 73
Omega Vehicle Security FPU5-MT-110, FPU5-MT-220 manual Specifications, 1.Calibration Accuracy