Troubleshooting Guide

7

 

 

Error Message

Problem

Solution

OVERCURRENT!

(continued)

6.Make sure the tubing is not caught on the bottom roller. Reposition the roller slightly and reload the tubing, if necessary.

7.Make sure no part of the tubing is being constricted by a clamp. Loosen clamps if necessary.

8.Turn the unit’s power off and on, adjust the settings, and attempt to resume pumping.

Select a MODE!

You have attempted to start the

 

 

pump, either while the display

 

shows the TUBE ID setting, or

 

after you had entered the

 

volume dispensed in the

 

calibration process.

1.Select a pumping mode (RPM, FLOW RATE, TIME, or TOTAL VOLUME).

2.Press the key that corresponds to that mode.

3.Use the UP and DOWN ARROW keys to adjust the setting.

4.Press the START/STOP key to start the pump.

Set RPM or RATE!

You have selected a command

 

 

that requires an RPM or FLOW

 

RATE setting to function properly.

1.Press the RPM or FLOW RATE key.

2.Use the UP and DOWN ARROW keys to adjust the setting.

3.Re-enter the original command and proceed.

SET TOTAL VOLUME

You have attempted to start the

 

 

pump in the TOTAL VOLUME

 

Mode with the TOTAL VOLUME

 

set at 0 mL.

1.Press the TOTAL VOLUME key

2.Use the UP and DOWN ARROW keys to adjust the TOTAL VOLUME setting.

3.Press the START/STOP key to start the pump.

7-5

Page 64
Image 64