Omega Vehicle Security FPU5-MT-110 4cont’d, Total Volume Mode, Display, Action Taken, Shows, 6-20

Models: FPU5-MT-110 FPU5-MT-220

1 83
Download 83 pages 6.26 Kb
Page 57
Image 57
Table 6-4 (cont’d)

 

 

Table 6-4 (cont’d)

TOTAL VOLUME MODE

 

 

KEY

 

DISPLAY

ACTION TAKEN

 

 

 

 

 

SHOWS

AND PROCEDURE

 

 

 

 

 

 

 

 

 

 

 

 

xxx mL/min

Displays actual FLOW RATE

 

 

FLOW

 

 

 

 

 

or

 

 

 

RATE

 

 

Press the FLOW RATE key. The display flashes the calculated

 

 

 

 

 

xxx L/min

 

 

 

 

 

 

 

 

 

 

 

fluid flow rate (xxx mL/min or xxx L/min).

 

 

 

 

 

 

 

 

 

 

 

 

xxx mL

Displays current TOTAL VOLUME setting

 

 

 

TOTAL

 

 

 

 

 

 

 

 

 

VOLUME

 

or

 

 

 

 

 

 

1. Press the TOTAL VOLUME key. The display flashes the

 

 

 

 

 

 

 

 

 

 

xxx Liters

 

 

 

 

 

TOTAL VOLUME setting you selected just before starting

 

 

 

 

 

 

 

 

 

 

 

 

the pump.

2.Use the UP and DOWN ARROW keys to enter the TOTAL VOLUME setting you desire. As you use the UP and DOWN ARROW keys, the display changes accordingly.

 

 

TUBE ID=xxx

Displays current TUBE ID

TUBE ID

 

 

 

 

 

 

Press the TUBE ID key. If the red LED next to the CALIBRATE

 

 

 

 

 

 

key is not lit, the display flashes the current TUBE ID setting. If

 

 

 

the red LED next to the CALIBRATE key is lit, the display flashes

 

 

 

the message “CALIBRATED TUBE!”.

 

 

CALIBRATED!

Displays calibration state

 

CALIBRATE

 

or

 

 

 

 

 

 

not calibrated

1. Press the CALIBRATE key. If the red LED next to the

 

 

 

 

 

CALIBRATE key is not lit, the display shows “not

 

 

 

calibrated”. If the red LED next to the CALIBRATE

 

 

 

key is lit, the display shows “CALIBRATED!”.

 

 

 

To calibrate a tube:

2.Press the START/STOP key to stop the pump. The display returns to the STOPPED Display Mode. For detailed instructions on “Calibrating a Tube”, refer to Section 5.5.

 and 

xxx mL

NOTE: if you press the UP and DOWN ARROW keys, the display

shows the current TOTAL VOLUME setting. This setting

 

or

 

xxx Liters

changes accordingly as you use the UP and DOWN

 

 

ARROW keys.

 

 

 

Language

 

NOTE: if you press the Language button, the display changes

Button

 

to the next language in the sequence: English, French,

 

 

Spanish, German, and Italian.

6-20

Page 57
Image 57
Omega Vehicle Security FPU5-MT-110 manual 4cont’d, Total Volume Mode, Display, Action Taken, Shows, And Procedure, 6-20