Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

4.4.2Multi-Point Calibration Summary

Following is a simple step-by-step procedure for executing a “multi-point” calibration of a Model 9816 Intelligent Pressure Scanner. It is assumed that all channels in the unit have the same full-scale pressure range. Optional commands are shown within brackets [ ]. If it should become necessary to abandon this calibration procedure once it is started, you may execute the Abort sub-command [C 03] of ‘C’ at any time after the first ‘C’ sub-command.

Description

TCP/IP Data

 

 

Insure that valves in RUN/CAL mode (default).

[w1200]

 

 

 

 

 

 

 

 

 

 

… normal data acquisition assumed to be running

 

 

 

 

 

 

 

 

 

 

 

Place the module calibration manifolds into the CAL

[w0C01] for CAL pos.

 

 

 

 

position if this is the desired pressure application input.

 

 

 

 

 

The RUN position may be a better choice for modules

[w0C00] for RUN pos.

 

 

 

 

with transducers having different ranges.

 

 

 

 

 

Ready the module for multi-point calibration by executing

C 00 FFFF 3 1 64

 

 

 

 

the Configure & Start (‘00’) sub-command of ‘C’. This

 

 

 

 

 

establishes all the channels to be affected, and

 

 

 

 

 

determines the total number of calibration points that will

 

 

 

 

 

be supplied (3 in this example) in later steps. It also

 

 

 

 

 

starts module averaging for calibration (64 samples in this

 

 

 

 

 

example). The linear fit (1) is required.

 

 

 

 

 

Apply 1st calibration pressure to the module’s CAL or

[rFFFF0]

 

 

 

 

RUN inputs. The zero (0.0) point is assumed in this case.

 

 

 

 

 

After applying zero pressure verify that this pressure is

 

 

 

 

 

measured correctly by the module.

 

 

 

 

 

When the data are stable, enter the Collect Data (‘01’)

C 01 1 0.0

 

 

 

 

sub-command of ‘C’ specifying this first calibration point

 

 

 

 

 

(1) with zero pressure (0.0).

 

 

 

 

 

Apply 2nd calibration pressure to the module’s CAL or

[rFFFF0]

 

 

 

 

RUN inputs. A full-scale (+5 psi) point is assumed in this

 

 

 

 

 

case. Verify that pressure reads correctly.

 

 

 

 

 

When the data are stable, enter another Collect Data

C 01 2 5.0

 

 

 

 

(‘01’) sub-command of ‘C’ specifying this second

 

 

 

 

 

calibration point (2) with 5.0 psi pressure.

 

 

 

 

 

Apply 3rd calibration pressure to the module’s CAL or

[rFFFF0]

 

 

 

 

RUN inputs. A mid-scale negative (-2.5 psi) point is

 

 

 

 

 

assumed in this case. A vacuum pump is normally

 

 

 

 

 

required to achieve such a pressure with 903x calibrators.

 

 

 

 

 

Verify that measured pressure reads correctly.

 

 

 

 

 

When the data are stable, enter last Collect Data (‘01’)

C 01 3 -2.5

 

 

 

 

sub-command of ‘C’ for this point (3) with a negative

 

 

 

 

 

(-2.5 psi) pressure.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 108

www.PressureSystems.com

Page 116
Image 116
Pressure Systems 9816, 98RK user manual Multi-Point Calibration Summary