READ TRANSDUCER TEMPERATURE (Command 't')

Purpose: Returns the most recently acquired engineering-unittemperature data (in ºC) for the specified channels. Each datum returned in the response will be in the specified high-precision data format.

Command t[p]ppppf

 

 

‘t’

is the command letter

 

 

 

 

 

‘[p]pppp’ is the

position field

 

 

 

 

 

f’ is the format field

 

 

 

 

 

 

 

 

 

Response

dddd[ dddd]...

 

 

 

 

 

 

 

 

 

 

 

 

dddd’ are datum fields, each with a leading space (except f =7).

 

 

 

 

 

 

 

Description:

 

The 4 or 5 character hex position field ([p]pppp) specifies a 16 or 20 bit

 

 

 

binary bit-map, with each bit set to 1 to specify a particular channel

 

 

 

number (S & P external rack channels, plus internal channels 16-1, left-

 

 

 

to-right).

 

 

 

 

 

 

 

The 1-character format field (f) specifies the format of each data field

 

 

 

(dddd) that will be returned in the requested response. The first datum

 

 

 

returned in the response will be for the highest channel number

 

 

 

specified. Each datum will be preceded by a space character. Some

 

 

 

formats may not be applicable to the specific type of data being

 

 

 

requested. Valid formats are shown in the following table:

 

 

 

 

 

 

 

 

f

 

Converts each internal response datum value from..

max.char.

 

0

single binary float

to

7-10-digit signed decimal “ [-xxx]x.xxxxxx”

13

 

 

 

 

 

 

 

 

 

1

single binary float

to

8-digit hex integer “ xxxxxxxx”

9

 

 

 

 

 

 

 

 

 

2

double binary float

to

16-digit hex integer “ xxxxxxxxxxxxxxxx”

17

 

 

 

 

 

 

 

 

 

5

single binary float

to

long integer (EU*1000) then to 8-digit hex

9

 

 

 

 

 

 

 

integer

 

 

 

7

single binary float

to

single binary float (big endian: msb first)

4

 

 

 

 

 

 

 

 

 

8

single binary float

to

single binary float (little endian: lsb first)

4

 

 

 

 

 

 

 

 

 

 

 

Example:

Send TCP/IP command to 9816 module (via its connected socket) that returns decimal temperature data for channels 1, 5, 9, and 13:

t11110

Response contains data (in ￿C) for channels 13, 9, 5, and 1 (left to right):

"21.234000 20.989500 21.005390 20.899602"

Page 85

www.PressureSystems.com

Page 93