Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Command 'C'— Sub-command Index 03: Abort Multi-Point Calibration

This sub-command has no additional parameters.

Command

C 03

 

 

‘C’ is the command letter.

 

‘ 03’ is the sub-command index (ii) for Abort.

 

NOTE: all parameters are separated by a space.

 

 

Response

A

 

 

‘A’ is the acknowledge letter

 

 

Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This sub-command also restores the original “A/D samples for averaging” count to the module that was in use before the calibration function was started.

It should be noted that executing the Configure & Start (‘00’) sub-command again, after the calibration function has started collecting data (per Collect Data (‘01’) sub-commands), but before the final data are calculated (per Calculate & Apply (‘02’) sub-command), will have the same affect as this Abort function.

Example:

Abort the Multi-Point Calibration function previously started “C 03

Read response:

"A"

Page 51

www.PressureSystems.com

Page 59