Pressure Systems 9816, 98RK user manual Zero and Gain Calibration Troubleshooting

Models: 98RK 9816

1 159
Download 159 pages 52.35 Kb
Page 150
Image 150

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

computer and the NetScannerSystem module are both properly configured for TCP/IP communications.

If an error free ping reply was not received, rerun the ping application using the IP address of the host computer. This will verify if the TCP/IP protocol was properly configured on the host computer. If a ping reply was not received, verify the TCP/IP installation steps for your host computer. Also verify that the host computer is configured for the proper IP address and subnet mask.

If the ping test of the host computer’s IP passed while the ping of the NetScannerSystem module failed, check the following possible sources for error:

Ensure the NetScannerSystem module’s IP has been assigned (as explained in Section 6.1.3.1) and the correct IP was used for the ping test.

Ensure the IP address of the host computer and the NetScannerSystem module are not duplicated on the network.

Ensure the link (LNK) LEDs are active on both the 9816 and the 98RK-1 for the installed slot. Also ensure the link LEDs are active on the host computer’s Ethernet adapter and the optional hub or switch to which it is attached as well as the 98RK-1 Host Link LED.

Ensure the Ethernet adapter card installed in the host is properly configured without conflict. In Windows®, this can be verified by entering the CONTROL PANEL under SETTINGS. Under CONTROL PANEL select the SYSTEM icon. When the DEVICE MANAGER tab is selected, a list of all installed hardware devices will be listed. Any possible hardware conflicts will be marked in this list with a yellow warning symbol next to the device in question.

Ensure the Ethernet adapter is configured for 10 Mbit/Sec. Many adapters are capable of higher speeds that are not compatible with the NetScannerSystem modules.

6.2Zero and Gain Calibration Troubleshooting

Incorrect pneumatic setup or incorrect command usage when executing a module’s Re-zero or Span calibration command (see 'Z', ‘h’, and 'C' commands in Chapter 3) can result in unexpected module operation. A common source of errors during these operations is incorrect control of the module’s internal calibration valve and pneumatic inputs.

Pressure connections are described in Chapter 2 while details of calibration procedures are described throughout Chapter 4. Some common errors and problems are listed below. These common problems apply primarily to Model 9816, with its internal transducers and calibration manifold.

The module’s supply air is either not attached or does not provide enough pressure (less than 65 psig) to shift the calibration valve. This results in the calibration valve remaining in its current position even though the module commands have requested movement of the valve. This causes incorrect pneumatic inputs during calibration commands. Verify this by reading the

Page 142

www.PressureSystems.com

Page 150
Image 150
Pressure Systems 9816, 98RK user manual Zero and Gain Calibration Troubleshooting