Pressure Systems 9816, 98RK user manual Example, 00 F 3 1 64 Read response

Models: 98RK 9816

1 159
Download 159 pages 52.35 Kb
Page 54
Image 54

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Example:

Configure and start the Multi-Point Calibration function so that it affects only the first four (4) channels of the module. Three (3) pressure calibration points will by supplied when we continue this function later (see example for ‘01’ sub-command below). A linear (1st order) fit will be used to obtain a new set of offset and gain correction coefficients for these four (4) channels. The maximum average sample count (64) is used to collect each calibration data point, so as to minimize any noise in the data samples. The module’s data acquisition process is altered immediately to collect the increased number of averages.

C 00 F 3 1 64” Read response:

"A"

Page 46

www.PressureSystems.com

Page 54
Image 54
Pressure Systems 9816, 98RK user manual Example, 00 F 3 1 64 Read response