5.1.6.3Adapter Plate O-Ring Replacement

Following is a step-by-step procedure to replace an adapter plate O-ring in a Model 9816 Intelligent Pressure Scanner. The adapter plate is located opposite the tubing plate on the calibration manifold. All DH200 transducers are attached to the adapter plate.

(1)Disassemble the module as described in Section 5.1.2.

(2)Remove the PC-206 Amplifier/Multiplexer board as described in Section 5.1.3.1. Lay the circuit board to the side on an anti-static surface.

(3)Remove the six (6) 3/32" Allen-head screws that secure the adapter plate to the calibration valve housing. To remove the two (2) center screws, you must remove the DH200 transducers near the screws. Make sure to note the DH200 serial number and location.

(4)Carefully rotate or slide the adapter plate back and forth, pivoting on the guiding pin about 1/8" several times. This is done to loosen the O-rings from the calibration manifold. Lift the adapter plate straight up. Do not touch the calibration manifold.

(5)Remove and replace the O-rings needing maintenance using the procedure described in Section 5.1.6.1. Note that the O-ring seals use an additional Teflon cup seal placed on top of the O-ring. These Teflon seals do not require Krytox® grease.

(6)Examine the adapter plate and calibration valve surface to verify that no contaminants are on either surface. This generally requires microscopic examination. Replace the adapter plate by slowly placing the plate on the calibration manifold housing. Make sure that the O-ring is down toward the pneumatic sliding manifold and the guiding pin on the adapter plate fits into the mating hole of the calibration valve housing. Fasten the adapter plate evenly on all sides.

(7)Install the DH200 transducers that were previously removed. It is suggested to install them back in their original location.

(8)Replace the PC-206 board as described in Section 5.1.3.1 and reassemble the module.

(9)Test your scanner to ensure proper operation.

Page 126

www.PressureSystems.com

Page 134
Image 134
Pressure Systems 9816, 98RK user manual Adapter Plate O-Ring Replacement