DOWNLOAD INTERNAL COEFFICIENTS (Command 'v')

Purpose: Downloads one or more internal coefficients to the module.

Command

vfaacc[-cc]dddd[ dddd]...

 

 

‘v’ is the command letter.

 

f’ is the format field.

 

aa’ is the array index field.

 

cc[-cc]’ is coefficient index [or contiguous range].

 

dddd’ are the datum fields, each with a leading space.

Response

A

 

‘A’ is the acknowledge letter.

 

 

Description: The 1-character format field (f) is a single digit that defines the format of each coefficient to be downloaded in the command’s datum ( dddd) fields, with each datum preceded by a space character. Most coefficients have a floating point datum type (f=0-1), while others have an integer datum type (f=5). Sending a datum in the improper format will result in an “N08” error response. Valid format types are shown in the following table:

f

converts each datum parameter value (‘ dddd’) from..

max.char.

0

1-10 digit signed decimal “ [-xxx]x.[xxxxxx]”

to

single

13

 

 

 

binary float

 

1

8-digit hex “ xxxxxxxx”

to

single

9

 

 

 

binary float

 

5

8-digit hex “ xxxxxxxx”

to

long binary

9

 

 

 

integer

 

The 2-character array index field (aa) is a hexadecimal value selecting a particular internal coefficient array to receive the downloaded data. The first array index (aa=01) refers to channel 1's transducer, the 16th (aa=10) refers to channel 16's transducer. Finally, the last array (aa=11) refers to a special global array.

A2-character coefficient index field (cc) is a hexadecimal value that selects a particular coefficient within the specified array. Multiple contiguous coefficients may be specified by using a coefficient index “range” specified by adding a hyphen (negative sign) between two such indexes (cc-cc).

Transducer Coefficient arrays for internal channels 1-16 are selected with array indexes aa=01 through aa=10 (hex). A special single Other Coefficient array is selected with array index aa=11 (hex). All the valid coefficient (cc) indexes (for each of these arrays) are listed in the appropriate tables included in the description of the previous Read Internal Coefficients ('u') command.

Page 90