Pressure Systems 98RK, 9816 user manual V00800-01 0.000 1.000 Response returned is

Models: 98RK 9816

1 159
Download 159 pages 52.35 Kb
Page 99
Image 99

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Example:

Send TCP/IP command to 9816 module (i.e., via its connected socket) with replacement values for the channel’s offset and gain correction terms loaded into the module’s volatile memory (cc = 00-01). Load these into channel # 8's Transducer Coefficient array (aa=08):

v00800-01 0.000 1.000” Response returned is:

"A"

Send command to 9816 module (via its connected socket) to change its default EU output from psi to kPa. This will be done by changing the EU Pressure Conversion Scaler to 6.894757.

"V01101 6.894757" Response returned is

"A"

Page 91

www.PressureSystems.com

Page 99
Image 99
Pressure Systems 98RK, 9816 user manual V00800-01 0.000 1.000 Response returned is, V01101 6.894757 Response returned is