Pressure Systems 9816, 98RK user manual Calculate and SET Offsets Command h, Hpppp

Models: 98RK 9816

1 159
Download 159 pages 52.35 Kb
Page 82
Image 82

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

CALCULATE AND SET OFFSETS (Command 'h')

Purpose: Instructs a module to calculate new offset coefficients, with zero differential pressure (or a specified “generated” pressure) applied to the specified internal channels. These new coefficients update part of the module’s internal calibration coefficient database, used to convert any subsequent raw data into engineering units data. The new offset values are also returned in the response. This command is sometimes called a Re-zerocalibration.

Command

hpppp[ vv.vvvv]

 

 

‘h’ is the command letter

 

pppp’ is the position field

 

vv.vvvv’ is an [optional] pressure value preceded by a space

 

character.

 

 

Response

g.gggg[ g.gggg]...

 

 

g.gggg’ are the actual offset data values returned, each preceded by

 

a space.

Description: The position field may have 0 or 4 characters. If no position field is specified, offset coefficients for all internal module channels (16-1) will be calculated and returned. If a position field is specified, offset coefficients for only the channels whose bits are set (=1) will be calculated and returned. If the optional pressure value [ vv.vvvv] is specified (rarely used), the position field must be 4 characters, even when all channels are to be specified. Offset values are returned in the response in order of highest specified channel to lowest specified channel, with data formatted per an implied decimal format (f=0).

NOTE: This command may only be applied to the sixteen (16) internal channels. It is not applicable for the external 98RK-1’s purge/supply (P & S) channels.

Before acquiring data with this command, any addressed Model 9816 module will normally attempt to place the calibration valve in the CAL position, so that a zero differential pressure can be applied to all channels via the module’s CAL and CAL Ref input port. Simply leaving these ports unattached will allow the transducers to read the appropriate zero differential pressure if ambient air pressure is stable. After data are acquired, the calibration valve will be placed in the RUN position. To disable the automatic shifting of the calibration valve, refer to the Set Operating Options ('w') command (index 0B). The reader is also referred to Section 4.2 of Chapter 4 for additional details concerning the performance of a Re-zero Calibration.

The calculated offset values from the latest 'h' command will be lost Note when the module is powered off. To save these offset terms to each transducer’s nonvolatile memory refer to the Set Operating Options

('w') command (index 08).

Page 74

www.PressureSystems.com

Page 82
Image 82
Pressure Systems 9816, 98RK user manual Calculate and SET Offsets Command h, Hpppp