INSUFFICIENT PRESSURE SUPPORT ALARM

WILL STOP WHEN:

the mask leak has been rectified.

NOTE The High Leak alarm can be turned off in the Leak Alert option in the Settings menu.

INSUFFICIENT PRESSURE SUPPORT ALARM

TRIGGERED BY:

air pressure at the mask has failed to reach an expected level for 3 breaths

the Air Circuit Learn function (page 26) has not been run after adding a new component (eg new mask, humidifier) to the system.

IF AN INSUFFICIENT PRESSURE SUPPORT ALARM SOUNDS:

check that the pressure sensor tubing is not kinked

check that the pressure sensor tubing is properly connected

run LEARN CIRCUIT.

WILL STOP WHEN:

the pressure sensor tubing is unobstructed and properly connected to the flow generator and mask

treatment is stopped (Start/Stop key or SMARTSTOP)

the AUTOSET CS2 is turned off.

FLOW BLOCKED ALARM

TRIGGERED BY:

blockage in air circuit

IF A FLOW BLOCKED ALARM SOUNDS:

1.The treatment will stop.

2.Turn power off.

3.Check whether there is a blockage in the air circuit.

4.Remove blockage.

5.Turn power back on.

6.If the Flow blocked alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal.

WILL STOP WHEN:

blockage is removed.

ENGLISH

THE ALARMS

33