TECHNICAL SPECIFICATIONS
DIMENSIONS (H X W X D)
142 x 248 x 293mm (5.6" x 9.8" x 11.5")
WEIGHT
Flow generator with cover: 3.7kg
Flow generator with HUMIDAIRE 2I (empty): 4.2kg
ENVIRONMENTAL CONDITIONS
Operating temperature: +5oC to +40oC
Storage and transport temperature:
Humidity — operating, storage and transportation:
HOUSING CONSTRUCTION
Injection moulded plastic
AIR TUBING
1 x 2m
AIR FILTER
Synthetic fibre
PATIENT CONNECTION PORT
22mm taper, compatible with EN
POWER SUPPLY
AC Input range:
DC: 30V (via ResMed DC/DC converter)
PERFORMANCE
Operating pressure range:
SOUND PRESSURE LEVEL
<30 dB (tested in accordance with the requirements of ISO
BATTERY TYPE
Lithium
ALARM SPECIFICATIONS
All alarms comply with EN 475 for Medium priority alarms: 3 beeps (935 Hz) repeated at intervals of 25 seconds; yellow LED flashes.
•HIGH MAXIMUM MASK PRESSURE ALARM
Stops air delivery at 25 cmH2O >700 msec.
•HIGH MEAN MASK PRESSURE ALARM
Stops air delivery when mean mask pressure exceeds 15 cmH2O (averaged over approx 1 minute).
•LOW MASK PRESSURE ALARM
Mask pressure is less than 3 cmH2O for longer than 5 seconds while the patient is using the AUTOSET CS2.
TECHNICAL SPECIFICATIONS
ENGLISH
43