ResMed AutoSet CS 2 user manual Leak exceeds 30 L/min for more than 20 seconds

Models: AutoSet CS 2

1 112
Download 112 pages 14.01 Kb
Page 48
Image 48
Leak exceeds 30 L/min for more than 20 seconds.

INSUFFICIENT PRESSURE SUPPORT ALARM

Is triggered if for 3 breaths in a row the maximum mask pressure achieved at inspiratory to expiratory transition is either <87.5% of the absolute mask pressure targetted at that point, or < absolute mask pressure targetted minus 1 cmH2O, whichever is the lower.

HIGH LEAK ALARM

Leak exceeds 30 L/min for more than 20 seconds.

FLOW BLOCKED ALARM

Flow is less than 6 L/min for more than 20 seconds.

IEC 601-1CLASSIFICATIONS

Class II, Type CF

Class II medical electrical equipment provides protection against electric shock either by double insulation or reinforced insulation and does not require a protective earth.

Table 3: Displayed values

Value

Range

Accuracy

Display

Resolution

 

 

 

 

 

 

 

Proximal cuff

 

 

 

 

 

 

 

Pressure

-3 to 40 cmH2O

0.5 cmH2O

0.1 cmH2O

 

 

 

 

Mass flow sensor

 

 

 

 

 

 

 

Leak

0–99 litres per

+/- 12 litres per

1 litre per breath

 

minute

minute

 

 

 

 

 

Respiratory rate

8–30 breaths per

+/- 10%

1 breath per

 

minute

 

minute

 

 

 

 

Tidal volume

0–4500 millilitres

+/- 10%

1 millilitre

 

 

 

 

Minute ventilation

0–40 litres per

+/- 10%

0.1 litres per

 

minute

 

minute

 

 

 

 

Target MV

0–40 litres per

+/- 10%

0.1 litres per

 

minute

 

minute

 

 

 

 

Table 4: Maximum low impedance flow at stated pressures

Pressure

Flow (L/min)

(cmH2O)

 

 

 

4.2217

8.6218

13.0219

44

Page 48
Image 48