ResMed AutoSet CS 2 user manual Guidance and manufacturer’s declaration - electromagnetic immunity

Models: AutoSet CS 2

1 112
Download 112 pages 14.01 Kb
Page 52
Image 52
Guidance and manufacturer’s declaration – electromagnetic immunity

Guidance and manufacturer’s declaration – electromagnetic immunity

The AutoSet CS2 is intended for use in the electromagnetic environment specified below. The customer or the user of the AutoSet CS2 should assure that it is used in such an environment.

Immunity test

IEC60601-1-2

Compliance level

Electromagnetic

test level

environment –guidance

 

 

 

 

 

 

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood,

discharge (ESD)

 

 

concrete or ceramic tile. If

IEC 61000-4-2

±8 kV air

±8 kV air

floors are covered with

 

 

 

synthetic material, the

 

 

 

relative humidity should be

 

 

 

at least 30%.

 

 

 

 

Electrical fast

±2 kV for power

±2 kV

Mains power quality should

transient/burst

supply lines

 

be that of a typical

IEC 61000-4-4

 

 

commercial or hospital

 

±1 kV for input/

Not Applicable

environment.

 

output lines

 

 

 

 

 

 

Surge

±1 kV differential

±1 kV differential

Mains power quality should

IEC 61000-4-5

mode

mode

be that of a typical

 

 

 

commercial or hospital

 

±2 kV common

±2 kV common

environment.

 

mode

mode

 

 

 

 

 

Voltage dips, short

<5% Ut

< 12V

Mains power quality should

interruptions and

(>95% dip in Ut)

(>95% dip in 240V)

be that of a typical

voltage variations on

for 0.5 cycle

for 0.5 cycle

commercial or hospital

power supply input

 

 

environment.

lines.

40% Ut

96V

If the user of the AutoSet CS2

IEC 61000-4-11

(60% dip in Ut)

(60% dip in 240V)

requires continued operation

 

for 5 cycles

for 5 cycles

during power mains

 

 

 

interruptions, it is

 

70% Ut

168V

recommended that the

 

(30% dip in Ut)

(30% dip in 240V)

AutoSet CS2 be powered

 

for 25 cycles

for 25 cycles

from an uninterruptible

 

<5% Ut

<12V

power source

 

 

 

(>95%dip in Ut)

(>95%dip in 240V)

 

 

for 5 sec

for 5 sec

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic

(50/60 Hz)

 

 

fields should be at levels

magnetic field

 

 

characteristic of a typical

IEC 61000-4-8

 

 

location in a typical

 

 

 

commercial or hospital

 

 

 

environment

 

 

 

 

NOTE: Ut is the a.c. mains voltage prior to application of the test level.

 

 

 

 

 