Guidance and manufacturer’s declaration – electromagnetic immunity (continued)

The AutoSet CS2 is intended for use in the electromagnetic environment specified below. The customer or the user of the AutoSet CS2 should assure that it is used in such an environment.

Immunity test

IEC60601-1-2

Compliance

Electromagnetic environment – guidance

test level

level

 

 

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any part

 

 

 

of the AutoSet CS2, including cables, than the

 

 

 

recommended separation distance calculated

 

 

 

from the equation applicable to the frequency of

 

 

 

the transmitter.

 

 

 

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

d = 1.17 √P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

Radiated RF

10 V/m

10 V/m

d = 0.35 √P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

 

 

 

 

d = 0.70 √P 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating of

 

 

 

the transmitter in watts (W) according to the

 

 

 

transmitter manufacturer and d is the

 

 

 

recommended separation distance in meters (m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site survey,a

 

 

 

should be less than the compliance level in each

 

 

 

frequency range.b

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AutoSet CS2 is used exceeds the applicable RF compliance level above, the AutoSet CS2 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the AutoSet CS2.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

ENGLISH

TECHNICAL SPECIFICATIONS

49

Page 53
Image 53
ResMed AutoSet CS 2 user manual Recommended separation distance

AutoSet CS 2 specifications

ResMed AutoSet CS 2 is a state-of-the-art device designed for individuals suffering from complex sleep apnea and central sleep apnea. This advanced treatment solution leverages innovation and cutting-edge technology to provide optimal comfort and effective therapy during sleep, ultimately helping users achieve a better quality of life.

One of the standout features of the AutoSet CS 2 is its sophisticated auto-adjusting pressure capabilities. The device automatically detects and responds to the patient's breathing patterns throughout the night, adjusting the pressure levels in real time to ensure that airway obstructions are effectively managed. This technology helps prevent both hypopneas and apneas, providing a seamless therapy experience.

The device is equipped with ResMed's proprietary Easy-Breathe technology, which allows for smooth pressure changes that mimic the user's natural breathing cycle. This feature enhances user comfort, making it easier for individuals to adapt to CPAP therapy and continue using the device consistently over time.

Another remarkable characteristic of the AutoSet CS 2 is its integrated wireless connectivity. This connectivity facilitates remote monitoring and data sharing with healthcare professionals, enabling more effective management of treatment progress. By analyzing sleep data, providers can make informed adjustments to therapy, ensuring that the patient receives the best possible care.

The AutoSet CS 2 also includes a user-friendly interface with a high-resolution color display, making it easy for users to navigate menus and access vital information about their sleep patterns and therapy effectiveness. With its compact design, the device is portable and can easily be used at home or while traveling.

In terms of comfort, the AutoSet CS 2 features a quiet operation, enabling a peaceful sleep environment. It is compatible with various masks, providing flexibility for users to find the best fit for their personal comfort preferences.

Overall, the ResMed AutoSet CS 2 is a leading choice for individuals with complex sleep apnea. With its combination of advanced features, user-centric design, and state-of-the-art technology, it empowers users to take control of their sleep and improve their overall health and well-being.