TROUBLESHOOTING

TROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the flow generator.

Problem

Possible cause

Solution

 

 

 

Insufficient air delivered from

Air filter is dirty.

Replace air filter.

flow generator

 

 

 

Air tubing is kinked or

Straighten or replace tubing.

 

punctured.

 

 

 

 

Flow generator operating

Water may have entered the

Remove the water from the

erratically, ie, switching on and

air tubing.

air tubing.

off, pressure varying at random

 

 

 

 

 

Pressure rises inappropriately

You are talking, coughing,

Avoid talking with the mask

 

voluntarily breath-holding (eg

on, and breathe as normally

 

while rolling over in bed), or

as possible.

 

intentionally breathing in an

 

 

unusual manner.

 

 

Severe mask leak or mouth

Correct your leak condition.

 

leak (greater than 30 L/min).

 

 

 

 

Alarm stays on continuously

Blockage in air circuit.

Ensure there is no barrier to

 

 

the passage of air.

 

Internal malfunction.

Return the device to your

 

 

equipment supplier for

 

 

service.

 

 

 

No LCD display

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected, and that the

 

 

switch at the back of the unit

 

 

is in the “on” position.

 

 

 

FAULT message displayed on

Internal malfunction.

Return the device to your

LCD

 

equipment supplier for

 

 

service.

ENGLISH

System Error 7 message

Device power-on has been

 

interrupted by breathing into

 

the mask during warm up.

Turn the flow generator off and on again. Do not place mask on face during warm up (30 seconds). Select LEARN CIRCUIT (page 26). If the error is not cleared, contact your equipment supplier for service.

Internal malfunction.

Return the device to your

 

equipment supplier for

 

service.

 

 