Troubleshooting

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit.

Problem

Possible Cause

Solution

 

 

 

No display.

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected and that the

 

 

switch at the back of the

 

 

unit is in the ON position.

 

 

 

Insufficient air

Ramp Time is in use.

Wait for air pressure to

delivered from the

 

build up.

VPAP.

 

 

 

Air filter is dirty.

Replace air filter.

 

Air tubing is kinked or

Straighten or replace

 

punctured.

tubing.

 

Air tubing not connected

Check air tubing.

 

properly.

 

 

Mask and headgear not

Adjust position of mask and

 

positioned correctly.

headgear.

 

Plug(s) missing from access

Replace plug(s).

 

port(s) on mask.

 

 

Pressure required for

See your clinician to adjust

 

treatment may have

the pressure.

 

changed.

 

 

 

 

The VPAP does not

Power cord not connected

Connect power cord firmly

start when you breathe

properly.

at both ends.

into the mask.

 

 

 

Power outlet may be faulty.

Try another power outlet.

 

The VPAP unit not switched

Switch power switch at rear

 

on.

of the VPAP to ON.

 

SmartStart not on.

Enable SmartStart.

 

Mask Alarm has been

Disable Mask Alarm to

 

enabled; SmartStart has

enable SmartStart.

 

automatically been

 

 

disabled.

 

 

Breath is not deep enough

Take a deep breath in and

 

to trigger SmartStart.

out through the mask.