GAP [X.XXV]
INPUT [X.XV]
Section 3. Configuration
GAP - When setting the “gap” threshold, the voltage shown on the top line of the display must be measured with nothing but the backing in the sensor and then again with a label still attached to the backing. The formula to be used for setting the threshold is:
(High Voltage Level + Low Voltage Level) x 0.5 = Start Value
1.Insert a label still attached to the backing into the sensor and close the Label
2.Strip the label from the backing and insert the backing strip under the sensor and close the Label Lid. Record the voltage shown on the top line of the LCD panel. The voltage ranges measured should be within the following ranges:
Backing with label = 2.0V to 3.5V Backing without label = Less than 1.0V
If the measured values are outside this range, you may have trouble in finding a value that will work properly under all conditions. If this is the case, a higher quality label may be needed to get adequate performance.
3.Calculate the starting point voltage using the formula.
4.Use the LINE and FEED keys to step the counter to the desired setting. The reading will advance to a setting of 4.9 (the maximum voltage). If a value of “0.0” is set, the printer will automatically set the level each time the printer is powered on with labels loaded and the head is closed.
5.Once the setting is correct, pressing the ENTER key will accept the setting and advance the next display.
SATO | PN 9001105 Rev. A | Page |