3. Click the right mouse button, and then select Device Configuration from
the drop-down menu that is displayed.
Result: The 9000 Series Meter Configuration window of the selected
meter is displayed.
4. Click the Advanced button.
Result: The first level window for 9000 Series meter programming is
displayed. The window lists all the groups (types) of modules available
in the selected meter. WinPM displays each module type or group in
one folder. The folders are sorted in alphabetically order. Each group
consists of modules having similar functionality.
5. Double click a module group folder such as, SWinDemand Modules.
Result: The second level window for 9000 Series meter programming is
displayed. The window lists all the modules available in the selected
group.
6. Double click a module folder such as, SWinDemand 29 module.
Result: The third level window for 9000 Series meter programming is
displayed. The window shows the details of the selected module, such
as input registers, output registers and other modules connected to the
module.
7. Click the Select Not Connected button in the lower left-hand corner of
the window, and then double click the input register to be deleted.
Result: WinPM displays a warning message.
8. Click the Yes button.
Result: WinPM displays a confirmation message box.
9. Click the Yes button.
Result: The link is deleted.