Tandberg Data T40 manual Interpreting LEDs

Models: T40

1 95
Download 95 pages 21.36 Kb
Page 78
Image 78

RAS

Ticket

6034

6036

6039

01ZZ

0200

1001

1101-

1128

1129-

112C

Display Message

Drive n Tape Alert 34. Download Fail

Drive n Tape Alert

36.Drive

Temperature

Drive n Tape Alert

39.Diagnostics Required

Diagnostic number ZZ failed.

Initialization of the robotics failed.

Failed moving the robotics.

Failed to stow at slot n

Failed to stow at drive n

Description

interface.

The firmware download has failed because you have tried to use the incorrect firmware for this tape drive.

Environmental conditions inside the tape drive are outside the specified temperature range

The tape drive may have a fault.

The diagnostics number ZZ failed.

The robotics failed to complete the calibration or the inventory sequence.

The robotics was not able to move.

The robotics is not able to stow the cartridge to the magazine

Failed to insert cartridge in drive

Suggested Action

Obtain the correct firmware and try again.

Verify that the drive CRU fan is running.

Check for availability of diagnostic information and run extended diagnostics if applicable.

Check the tape drive users manual for instructions on running extended diagnostic tests and retrieving diagnostic data

Reboot the loader and retry the operation.

Reboot the loader and retry the operation.

Verify that the magazines are locked.

Turn off the loader and try to remove the magazines by using the emergency eject pin.

Contact your service representative.

Reboot the loader and retry the operation.

Reboot the loader and retry the operation.

Remove the magazine if possible. Remove all cartridges.

Verify that the cartridge is not damaged.

Insert empty magazines.

Power cycle the loader.

Retry the operation.

Retry the operation with a new magazine.

Reboot the loader and retry the operation.

If failure persists, insert the cartridge into a magazine slot, remove the magazine and verify that the cartridge is not damaged.

If failure persists, replace the drive CRU.

Interpreting LEDs

LEDs provide a visual indication about the status of certain library components. LEDs can sometimes communicate that a problem exists when the RAS ticket cannot. For example, an LED can indicate a firmware problem that prohibits the library from generating RAS tickets.

70