Model 360E

Instruction Manual Warranty/Repair
Questionnaire
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
TELEDYNE API CUSTOMER SERVICE
EMAIL: api-customerservice@teledyne.com
PHONE: (858) 657-9800 - TOLL FREE: (800) 324-5190 - FAX: (858) 657-9816
C-1 05235 Rev B
CUSTOMER: ____________________________________ PHONE: ______________________________________
CONTACT NAME: ________________________________ FAX NO: ______________________________________
SITE ADDRESS: __________________________________________________________________________________
SERIAL NO.: ____________________________________ FIRMWARE REVISION:__________________________
1. Are there any failure messages? ____________________________________________________________________
________________________________________________________________________________________________
________________________________________________________________________________________________
________________________________________________________________________________________________
________________________________________________________________________________________________
Please complete the following table:
PARAMETER DISPLAYED AS OBSERVED VALUE UNITS NOMINAL RANGE
Range PPM 0 -10, 0 - 2000 Ppm
Stability STABIL PPM 0.15 Ppm With Zero Air
CO2 Measure CO2 MEAS mV 3600 – 4800 Mv
CO2 Reference CO2 REF mV 1400 – 2000 Mv
Measure/Reference Ratio MR RATIO 2.5 ± 0.02 W/ Zero Air
Pressure PRES INHG -2”Ambient Absolute
Sample Flow SAMP FL CC 800 ± 10%
Sample Temp SAMPLE TEMP ºC 48 ± 4
Bench Temp BENCH TEMP ºC 48 ± 2
Wheel Temp WHEEL TEMP ºC 68 ± 2
Box Temp BOX TEMP ºC Ambient + 7 ± 10
Photo Drive PHT DRIVE mV 250 Mv TO 4750 Mv
Slope SLOPE . 1.0 ± .3
Offset OFFSET 0 ± 0.3
O2 Sensor Temperature3 O2 CELL TEMP
Slope of O2 Measurement3 O2 SLOPE
Offset of O2 Measurement3 O2 OFFSET
Dark Cal Reference signal REF DARK OFFSET mV 125 ± 50 Mv.
Dark Cal Measurement Signal MEAS DARK OFFSET mV 125 ± 50 Mv
Electric Test PPM 40 ± 2 Ppm
Values are in the Signal I/O
REF_4096_MV 4096mv±2mv And Must Be
Stable
REF_GND 0± 0.5 And Must Be Stable
2. Have you performed a leak check and flow check?______________________________________________________
3. What is the sample flow & sample pressure with the sample in-let on rear of machine capped?
SAMPLE FLOW_________________________ CC SAMPLE PRESSURE ______________________IN-HG-A