Teledyne API Model 465L O3 Monitor Operation Manual
8.5.3.Determining a Dynamically Assigned IP Address (DHCP Mode)
The DeviceInstaller software application (See Section 8.5) can be used to search a network for instruments and determine the IP Address assigned by a DHCP server.
1.Click
2.Click on the Search icon. After a moment a list of instruments on the network will be displayed. The M465L should be listed as an
3.
8.5.4.Manually Assigning a Static IP Address
For applications where the IP Address must remain constant, a static IP Address can be manually assigned to the instrument and the DHCP function must be disabled.
1.Click
Note: If the unit already has an IP address (e.g., DHCP has assigned an IP address), click the Search icon and select the unit from the list of Lantronix device servers on the local network.
2.Click the Assign IP icon.
3.If prompted, enter the hardware address (on the product label) and click Next.
4.Select Assign a specific IP address and click Next.
5.Enter the IP address. The Subnet mask displays automatically based on the IP address; if desired, you may change it. On a local network, you can leave the Default gateway blank (all zeros). Click Next.
6.Click the Assign button and wait several seconds until a confirmation message displays. Click Finish.
7.Select the device from the main window list and select Ping from the Tools menu. The Ping Device dialog box shows the IP address of the selected unit.
8.From the Tools menu, click the Ping button. The results display in the Status window. Click the Clear Status button to clear the window so you can ping the device again.
Note: If you do not receive “Reply” messages, make sure the unit is properly attached to the network and that the IP address assigned is valid for the particular network segment you are working with. If you are not sure, check with your systems administrator.
9.Click the Close button to close the dialog box and return to the main window.