7.2.6.SETUP-VARS Menu

The VARS menu allows viewing and editing of various global setup variables that effect how the M465L operates. These variables are stored in the instrument’s non-volatile memory. Table 7-7 lists these variables and their function.

Table 7-7: VARS List

 

VAR

 

 

Meaning

 

Default

 

Range

 

 

 

 

 

 

 

Value

 

 

 

 

S/N

 

 

Serial number of M465L. This number should match

N/A

 

1 – 9999

 

 

 

 

 

 

 

 

 

the serial number tag on the rear of the instrument.

 

 

 

 

 

ALARM

 

 

Controls the latching behavior of the concentration alarms.

 

LATCHING,

 

MODE

 

 

See Section 7.1.4 for additional details.

 

NON-LATCHING

 

STREAM

 

 

 

 

1

1,3,6

 

 

 

 

Specifies the number of streams available in the instrument. Note that the

 

 

COUNT*

 

 

hardware in the instrument must be also configured to support the number

 

 

 

 

 

 

 

 

of streams specified.

 

 

 

 

 

DWEL

 

 

Dwell value for stream selector operation. See Section 7.3

30.0

 

0.0 – 30.0

 

 

 

 

for additional details.

Seconds

 

Seconds

 

 

 

 

Multi-stream units with the Sample Conditioner option

45.0

 

0.0 – 45.0

 

 

 

 

(Section 4.1.1.2), require additional dwell time to purge the

Seconds

 

Seconds

 

 

 

 

filter between stream measurements.

 

 

 

 

 

ALARM

 

 

Number of readings above alarm limit required to trigger

5

1 - 10

 

 

DELAY

 

 

alarms

 

 

 

 

MAN MODE

 

 

Enables/disables the Manual mode timeout feature. See

ON

 

ON, OFF

TIMEOUT

 

 

Section 7.1.3 for more detail.

 

 

 

 

 

BC FILT

 

 

Boxcar filter length for concentration filter. Increasing the

10

1-30

 

 

SIZE*

 

 

size of the filter will smooth out the concentration readings

 

 

 

 

 

 

 

 

but will also increase the response time.

 

 

 

 

 

FLOW

 

 

Enables/disables operation of optional electronic flowmeter

OFF

ON, OFF

 

ENABLE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

COMM

 

 

MODBUS address for serial communications

1

1 - 20

 

 

ADDRESS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UNIT

 

 

Concentration unit-of-measure setting.

PPM

PPM, PPB, UGM

 

 

 

 

 

 

 

 

 

SCROLL

 

 

Inter-character delay for the front panel display, to increase

20

6-100

 

 

DELAY

 

 

speed (lower value) or decrease speed (higher value) the

 

 

 

 

 

 

 

 

display scroll.

 

 

 

 

USER SLOPE

 

User-editable calibration slope value

1.000

 

0.8 – 1.0

 

 

 

 

 

 

 

 

 

UNITS

 

 

Standard (20ºC) or Normal (0ºC) temperature selection for

NORMAL

 

NORMAL

 

TEMP

 

 

g/Nm3 units of measure.

 

 

STANDARD

* Changing the value of these VARS forces the instrument to automatically reboot for the changes to take effect.