Teledyne API Model 465L O3 Monitor Operation Manual
7.2.6.SETUP-VARS Menu
The VARS menu allows viewing and editing of various global setup variables that effect how the M465L operates. These variables are stored in the instrument’s
Table
| VAR |
|
| Meaning |
| Default |
| Range |
|
|
|
|
|
|
| Value |
|
|
|
| S/N |
|
| Serial number of M465L. This number should match | N/A |
| 1 – 9999 | ||
|
|
|
|
| |||||
|
|
|
| the serial number tag on the rear of the instrument. |
|
|
|
| |
| ALARM |
|
| Controls the latching behavior of the concentration alarms. |
| LATCHING, | |||
| MODE |
|
| See Section 7.1.4 for additional details. |
| ||||
| STREAM |
|
|
|
| 1 | 1,3,6 |
| |
|
|
| Specifies the number of streams available in the instrument. Note that the |
| |||||
| COUNT* |
|
| hardware in the instrument must be also configured to support the number |
|
|
|
| |
|
|
|
| of streams specified. |
|
|
|
| |
| DWEL |
|
| Dwell value for stream selector operation. See Section 7.3 | 30.0 |
| 0.0 – 30.0 | ||
|
|
|
| for additional details. | Seconds |
| Seconds | ||
|
|
|
| 45.0 |
| 0.0 – 45.0 | |||
|
|
|
| (Section 4.1.1.2), require additional dwell time to purge the | Seconds |
| Seconds | ||
|
|
|
| filter between stream measurements. |
|
|
|
| |
| ALARM |
|
| Number of readings above alarm limit required to trigger | 5 | 1 - 10 |
| ||
| DELAY |
|
| alarms |
|
|
|
| |
MAN MODE |
|
| Enables/disables the Manual mode timeout feature. See | ON |
| ON, OFF | |||
TIMEOUT |
|
| Section 7.1.3 for more detail. |
|
|
|
| ||
| BC FILT |
|
| Boxcar filter length for concentration filter. Increasing the | 10 |
| |||
| SIZE* |
|
| size of the filter will smooth out the concentration readings |
|
|
|
| |
|
|
|
| but will also increase the response time. |
|
|
|
| |
| FLOW |
|
| Enables/disables operation of optional electronic flowmeter | OFF | ON, OFF | |||
| ENABLE |
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| |
| COMM |
|
| MODBUS address for serial communications | 1 | 1 - 20 |
| ||
| ADDRESS |
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| |
| UNIT |
|
| Concentration | PPM | PPM, PPB, UGM | |||
|
|
|
|
|
|
|
| ||
| SCROLL |
|
| 20 |
| ||||
| DELAY |
|
| speed (lower value) or decrease speed (higher value) the |
|
|
|
| |
|
|
|
| display scroll. |
|
|
|
| |
USER SLOPE |
| 1.000 |
| 0.8 – 1.0 | |||||
|
|
|
|
|
|
|
| ||
| UNITS |
|
| Standard (20ºC) or Normal (0ºC) temperature selection for | NORMAL |
| NORMAL | ||
| TEMP |
|
| g/Nm3 units of measure. |
|
| STANDARD |
* Changing the value of these VARS forces the instrument to automatically reboot for the changes to take effect.