Teledyne API Model 465L O3 Monitor Operation Manual

10.2.Troubleshooting Using Front Panel Status LED’s or Status Outputs

The M465L has 4 front-panel status LED’s that reflect the current operating status of the monitor, and indicate fault conditions. There are also four relay Status Outputs on the rear panel (or internally in the NEMA configuration) that also reflect the state of these status LED’s. See Section 4.2.3 for more information on the Status Outputs.

The Status LED’s are categorized into Critical and Non-Critical warnings. Critical warnings are those that would normally require immediately removing the monitor from service and repairing it. Non-critical warnings are those that indicate some maintenance would be useful, but not immediately required.

Table 10-1: Status LED/Output Definitions

Status

Status LED

Normal

Triggers

Critical

Output #

Label

Operatin

 

Warning?

 

 

g State

 

 

1

Sensor OK

On

∙ No bench communications

Yes

 

 

 

∙ Reference < 125.0 mV

 

 

 

 

∙ Reference > 1230.0 mV

 

2

Invalid Reading

Off

∙ O3 Concentration < -10.0 PPB

No

 

 

 

∙ O3 Concentration > F.S. Range

 

3

Check Lamp

Off

∙ Reference < 250.0 mV

No

4

Pneumatic

Off

∙ Pressure < 9.0 psia

No

 

Error

 

∙ Pressure > 14.9 psia

 

 

 

 

∙ Flow < 640 cc/min

 

 

 

 

∙ Flow > 960 cc/min

 