Teledyne API Model 465L O3 Monitor Operation Manual
10.2.Troubleshooting Using Front Panel Status LED’s or Status Outputs
The M465L has 4
The Status LED’s are categorized into Critical and
Table 10-1: Status LED/Output Definitions
Status | Status LED | Normal | Triggers | Critical |
Output # | Label | Operatin |
| Warning? |
|
| g State |
|
|
1 | Sensor OK | On | ∙ No bench communications | Yes |
|
|
| ∙ Reference < 125.0 mV |
|
|
|
| ∙ Reference > 1230.0 mV |
|
2 | Invalid Reading | Off | ∙ O3 Concentration < | No |
|
|
| ∙ O3 Concentration > F.S. Range |
|
3 | Check Lamp | Off | ∙ Reference < 250.0 mV | No |
4 | Pneumatic | Off | ∙ Pressure < 9.0 psia | No |
| Error |
| ∙ Pressure > 14.9 psia |
|
|
|
| ∙ Flow < 640 cc/min |
|
|
|
| ∙ Flow > 960 cc/min |
|
80 | 05509F DCN6409 |