Teledyne API Model 465L O3 Monitor Operation Manual
4.2.4.Hi-Current Relay Outputs
Three form C relay outputs are provided on the rear panel on the nine pin connector (See Figure
The relays are dry contact type form C (SPDT) relays with Normally Open (N.O.), Normally Closed (N.C.) and Common (COM) contacts. The relays are capable of driving loads of up to 250VAC, 5A. For maximum contact life, the relays should only be used to drive resistive loads. Contact life may be dramatically shortened if inductive loads are driven without any provision for minimizing high voltage “inductive kick” that can occur.
These relays have slightly different functions depending on whether the monitor is a single stream or
Table
Relay # | Function |
| Normal |
|
|
| Operating State |
1 | System OK Diagnostic | On | |
2 | Hi Alarm | Off | |
3 | Off |
Table
Relay # | Function |
|
|
1System OK Diagnostic
2Global Hi Alarm*
3Global
Normal Operating State
On
Off
Off
*The state of the global alarm represents all the Hi or