Teledyne API Model 465L O3 Monitor Instruction Manual
Sample Conditioner Option
The Sample Conditioner option is a water
Due to the additional volume added by the water
The Response Time of an M465L instrument equipped with the Sample Conditioner option, will be <60 seconds to 95%.
CAUTION
The Sample Conditioner will NOT protect the unit from damage by large amounts of liquid, such as that ingested through immersion of the sample line.
Furthermore, it is important to closely track the amount of water accumulating in the Water
The Sample Conditioner option includes a Permeation Gas Dryer, which is available as an alternate option (described next) if humidity is the only concern and there is no risk of condensation.
Permeation Gas Dryer Option
The Permeation Gas Dryer serves to remove water vapor from sampling gas in a humid environment. It is an alternative option when there is no risk of condensation. The permeation material is a Teflon Type compound and has only a minimal effect on the ozone concentration in the sample stream. When the M465L is equipped with this option, the unit has been calibrated at the factory to compensate for any ozone loss in the permeation dryer.
05509F DCN6409 | 27 |