4.GETTING STARTED
The Model 465 is shipped with the following standard equipment:
∙
∙
Power cord (Rack mount version only) Instruction manual.
Upon receiving the Model 465 please do the following:
1.Verify that there is no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)
2.When installing the Model 465, allow a minimum of 4 inches for clearance at the back of the instrument (rack mount only) and 1 inch of clearance on each side for proper ventilation.
3.Connect sample inlet line(s) to the sample port on rear panel or bottom panel of
instrument. See Figure
CAUTION
Connect the exhaust fitting on the rear panel (See Figure
4.Ensure that there are no communication devices connected until after
5.For the NEMA configuration, the electrical connection must be
∙
∙
∙
NOTE
Sample tubing made from an inert material such as Teflon should
be used to minimize sample degradation.
Earth Ground (green): Connect the earth ground wire to the screw terminal lug on the chassis as shown in the Figure below. Do not connect the earth ground wire directly to the PCB connector on the mainboard PCB.
Line (Black): Connect the Line wire to the connector on the mainboard PCB labeled “Line”.
Neutral (White): Connect the Neutral wire to the connector on the mainboard PCB labeled “Neutral”.
05509F DCN6409 | 21 |