Teledyne API Model 465L O3 Monitor Operation Manual

8.6.3.Real-Time Concentration and Status Registers (Read Only)

Decimal

 

 

 

 

 

Address

Description

Register Type

Start Bit

# Bits

 

1

 

Stream 1 Concentration (PPB)

Float Inverse

0

32

 

3

 

Stream 2 Concentration (PPB)

Float Inverse

0

32

 

5

 

Stream 3 Concentration (PPB)

Float Inverse

0

32

 

7

 

Stream 4 Concentration (PPB)

Float Inverse

0

32

 

9

 

Stream 5 Concentration (PPB)

Float Inverse

0

32

 

11

 

Stream 6 Concentration (PPB)

Float Inverse

0

32

 

13

 

Stream 1 HI Alarm Status, 1= ON

Byte

0

8

 

13

 

Stream 2 HI Alarm Status, 1= ON

Byte

8

8

 

14

 

Stream 3 HI Alarm Status, 1= ON

Byte

0

8

 

14

 

Stream 4 HI Alarm Status, 1= ON

Byte

8

8

 

15

 

Stream 5 HI Alarm Status, 1= ON

Byte

0

8

 

15

 

Stream 6 HI Alarm Status, 1= ON

Byte

8

8

 

16

 

Stream 1 HI-HI Alarm Status, 1= ON

Byte

0

8

 

16

 

Stream 2 HI-HI Alarm Status, 1= ON

Byte

8

8

 

17

 

Stream 3 HI-HI Alarm Status, 1= ON

Byte

0

8

 

17

 

Stream 4 HI-HI Alarm Status, 1= ON

Byte

8

8

 

18

 

Stream 5 HI-HI Alarm Status, 1= ON

Byte

0

8

 

18

 

Stream 6 HI-HI Alarm Status, 1= ON

Byte

8

8

 

19

 

Stream 1 Active Status, 1= Sampling

Byte

0

8

 

19

 

Stream 2 Active Status, 1= Sampling

Byte

8

8

 

20

 

Stream 3 Active Status, 1= Sampling

Byte

0

8

 

20

 

Stream 4 Active Status, 1= Sampling

Byte

8

8

 

21

 

Stream 5 Active Status, 1= Sampling

Byte

0

8

 

21

 

Stream 6 Active Status, 1= Sampling

Byte

8

8

 

24

 

Gas Flow (cc/min)

Float Inverse

0

32

 

39

 

Sensor Reference Value (mV)

Float Inverse

0

32

 

41

 

Sensor Measure Value (mV)

Float Inverse

0

32

 

43

 

Sensor Pressure Value (psia)

Float Inverse

0

32

 

45

 

Sensor Gas Temperature (K)

Float Inverse

0

32

 

47

 

Sensor UV Lamp Temp (°C)

Float Inverse

0

32

 

 

 

 

 

 

 

 

62

 

 

 

05509F DCN6409

Page 64
Image 64
Teledyne 465L operation manual Real-Time Concentration and Status Registers Read Only