Teledyne API Model 465L O3 Monitor Operation Manual
8.6.3.Real-Time Concentration and Status Registers (Read Only)
Decimal |
|
|
|
|
| |
Address | Description | Register Type | Start Bit | # Bits |
| |
1 |
| Stream 1 Concentration (PPB) | Float Inverse | 0 | 32 |
|
3 |
| Stream 2 Concentration (PPB) | Float Inverse | 0 | 32 |
|
5 |
| Stream 3 Concentration (PPB) | Float Inverse | 0 | 32 |
|
7 |
| Stream 4 Concentration (PPB) | Float Inverse | 0 | 32 |
|
9 |
| Stream 5 Concentration (PPB) | Float Inverse | 0 | 32 |
|
11 |
| Stream 6 Concentration (PPB) | Float Inverse | 0 | 32 |
|
13 |
| Stream 1 HI Alarm Status, 1= ON | Byte | 0 | 8 |
|
13 |
| Stream 2 HI Alarm Status, 1= ON | Byte | 8 | 8 |
|
14 |
| Stream 3 HI Alarm Status, 1= ON | Byte | 0 | 8 |
|
14 |
| Stream 4 HI Alarm Status, 1= ON | Byte | 8 | 8 |
|
15 |
| Stream 5 HI Alarm Status, 1= ON | Byte | 0 | 8 |
|
15 |
| Stream 6 HI Alarm Status, 1= ON | Byte | 8 | 8 |
|
16 |
| Stream 1 | Byte | 0 | 8 |
|
16 |
| Stream 2 | Byte | 8 | 8 |
|
17 |
| Stream 3 | Byte | 0 | 8 |
|
17 |
| Stream 4 | Byte | 8 | 8 |
|
18 |
| Stream 5 | Byte | 0 | 8 |
|
18 |
| Stream 6 | Byte | 8 | 8 |
|
19 |
| Stream 1 Active Status, 1= Sampling | Byte | 0 | 8 |
|
19 |
| Stream 2 Active Status, 1= Sampling | Byte | 8 | 8 |
|
20 |
| Stream 3 Active Status, 1= Sampling | Byte | 0 | 8 |
|
20 |
| Stream 4 Active Status, 1= Sampling | Byte | 8 | 8 |
|
21 |
| Stream 5 Active Status, 1= Sampling | Byte | 0 | 8 |
|
21 |
| Stream 6 Active Status, 1= Sampling | Byte | 8 | 8 |
|
24 |
| Gas Flow (cc/min) | Float Inverse | 0 | 32 |
|
39 |
| Sensor Reference Value (mV) | Float Inverse | 0 | 32 |
|
41 |
| Sensor Measure Value (mV) | Float Inverse | 0 | 32 |
|
43 |
| Sensor Pressure Value (psia) | Float Inverse | 0 | 32 |
|
45 |
| Sensor Gas Temperature (K) | Float Inverse | 0 | 32 |
|
47 |
| Sensor UV Lamp Temp (°C) | Float Inverse | 0 | 32 |
|
|
|
|
|
|
|
|
62 |
|
|
| 05509F DCN6409 |