Teledyne API Model 465L O3 Monitor Operation Manual
7.2.4.VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in
Note that these values are updated in
Table
Parameter | Meaning |
MEAS | UV detector reading, measure cycle |
REF | UV detector reading, reference |
| cycle |
PRESSURE | Sample pressure |
LAMP | UV lamp temperature |
TEMP |
|
SAMPLE | Sample temperature |
TEMP |
|
FLOW | Sample Gas Flow Rate |
*After initial 20 minute
Units
mV mV
psia ºC
ºC
cc/min
Normal Range*
250 – 1230
250 – 1230
9.0– 14.9
50.0– 52.0
5.0
640 – 960
7.2.5.CONFIG Menu
The CONFIG menu displays specific information related to the instrument’s configuration. No changes are made within this menu.
Table
CONFIG | Meaning |
S/N | Serial number of M465L. (Should |
| match the serial number tag on the rear |
| panel of the instrument). |
COMM ADDRESS | Displays the MODBUS slave ID for |
| the instrument. |
STREAM COUNT | Displays the number of streams |
| configured. (Always 1 for instruments |
| without the |
REV | Revision of instrument firmware |
BREV | Revision of sensor firmware |
BOOT | Revision of bootloader firmware |