Teledyne API Model 465L O3 Monitor Operation Manual

7.2.4.VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in real-time. This can be useful for diagnosing various instrument or system problems.

Note that these values are updated in real-time as they are repeatedly scrolled on the display.

Table 7-5: VIEW Menu Parameters

Parameter

Meaning

MEAS

UV detector reading, measure cycle

REF

UV detector reading, reference

 

cycle

PRESSURE

Sample pressure

LAMP

UV lamp temperature

TEMP

 

SAMPLE

Sample temperature

TEMP

 

FLOW

Sample Gas Flow Rate

*After initial 20 minute warm-up period.

Units

mV mV

psia ºC

ºC

cc/min

Normal Range*

250 – 1230

250 – 1230

9.0– 14.9

50.0– 52.0

5.0–50.0

640 – 960

7.2.5.CONFIG Menu

The CONFIG menu displays specific information related to the instrument’s configuration. No changes are made within this menu.

Table 7-6: CONFIG Menu Parameters

CONFIG

Meaning

S/N

Serial number of M465L. (Should

 

match the serial number tag on the rear

 

panel of the instrument).

COMM ADDRESS

Displays the MODBUS slave ID for

 

the instrument.

STREAM COUNT

Displays the number of streams

 

configured. (Always 1 for instruments

 

without the Multi-stream Option).

REV

Revision of instrument firmware

BREV

Revision of sensor firmware

BOOT

Revision of bootloader firmware

05509F DCN6409

51

Page 53
Image 53
Teledyne 465L operation manual View Menu, Config Menu, Parameter Meaning, Units, Normal Range