Teledyne API Model 465L O3 Monitor Operation Manual

9.5.UV Lamp Replacement

CAUTION

GENERAL SAFETY HAZARD

Do not look at the UV lamp while the unit is operating. UV light can cause eye damage. Always use safety glasses made from UV blocking material. (Generic plastic glasses are not adequate).

WARNING - RISK OF ELECTRICAL SHOCK

HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO

THE INSTRUMENT!

1.Disconnect power from the M465L.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Loosen the two UV lamp setscrews on the UV lamp housing (see Figure 9-1 or Figure 9-2 above.)

4.Unplug the lamp power cord from the connector labeled P1 on the sensor module.

5.Loosen the four slotted captive screws attaching the sensor module to the chassis.

6.Carefully slide the lamp out of housing. The sensor module may need to be raised slightly to completely remove the lamp.

7.Install the new lamp, seating it in the lamp housing until it stops.

8.Re-attach the sensor module captive screws to the chassis.

9.Re-tighten the two UV lamp setscrews.

10.Plug the lamp power cord into P1 on the sensor module.

11.Reconnect power to the instrument and turn on power switch. Let instrument warm up for at least 20 minutes.

12.Perform UV lamp adjustment procedure per Section 9.4.

9.6.Sample Pump Replacement

1.Disconnect power from the M465L.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Locate sample pump assembly (See Figure 9-1.)

4.Cut off the clamps holding the tubing to the pump nipples and remove tubing. Note which tubing is connected to inlet and outlet.

70

05509F DCN6409

Page 72
Image 72
Teledyne 465L operation manual UV Lamp Replacement, Sample Pump Replacement