Teledyne API Model 465L O3 Monitor Operation Manual
9.7.Sensor Valve Replacement
1.Disconnect power from the M465L.
2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)
3.Locate sensor module assembly (See Figure
4.Unplug the
5.Remove the silver retainer clip from the top of the sensor valve. A pair of pliers may be used to slide off the retainer clip.
6.The valve coil can now be removed by sliding upwards.
7.Remove the two (2) mounting screws using a #2 Phillips screwdriver. See Figure
8.Remove valve body from sensor manifold.
9.Clean any residue or dirt off the surface of the manifold using a
10.Install the new valve by reversing steps
Figure 9-3: Sensor Detail