Teledyne API Model 465L O3 Monitor Operation Manual

7.2.7.SETUP-STREAM Menu

The STREAM menu allows the operator to view and edit properties of each measurement stream available. For each stream the following properties are available:

Table 7-8: STREAM Properties

Property

Meaning

Default

Range

 

 

Value

 

DURATION

Stream measurement duration for

60.0 Seconds

60.0 – 360.0

 

AUTO stream-selector operation.

 

Seconds

 

See Section 7.3 for additional

 

 

 

details.

 

 

HI ALARM

HI alarm trigger value

0.100 PPM

.010 – 9.999

 

 

 

PPM*

HIHI ALARM

HI-HI alarm trigger value

0.300 PPM

.010 – 9.999

 

 

 

PPM*

ON/OFF HI

HI alarm enable, enables HI alarm

ON

ON/OFF

ALARM

operation

 

 

ON/OFF HI HI

HI-HI alarm enable, enables HI-HI

ON

ON/OFF

ALARM

alarm operation

 

 

RANG

Concentration range for specified

1.0 PPM

1.0 – 500.0

 

stream. Used to scale the analog

 

PPM

 

output.

 

 

ENABLE

Allows the choice of viewing or

ON

ON-OFF

 

skipping each stream in a multi-

 

 

 

channel instrument when cycling

 

 

 

through the streams in Auto or

 

 

 

Manual modes. OFF = skip. At

 

 

 

least one stream must be set to ON

 

 

 

(enabled).

 

 

* For low concentration instrument ranges. For certain range configurations, the alarms may be set to higher values

05509F DCN6409

53

Page 55
Image 55
Teledyne 465L operation manual SETUP-STREAM Menu, Property Meaning Default Range Value