2.0 Installation

Infrared Gas Analyzer

Installation 2

 

 

 

 

 

 

 

 

 

2.0 Installation

Installation of the Model 7300A Infrared Gas Analyzer includes:

1.Unpacking

2.Mounting

3.Gas connections

4.Electrical connections

5.Testing the system.

2.1Unpacking the Analyzer/Inspection

The analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the analyzer and inspect it for damage. Immediately report any damage to the shipping agent.

2.2Installing and Connecting the Analyzer

The 7300A analyzer is a general purpose analyzer and as such is designed with (non-sealed) enclosures. It must be installed in an area where the ambient temperature is not permitted to drop below 32°F nor rise above 100°F. In areas outside these temperatures, auxillary heating/cooling must be supplied. The 7300A enclosure is oil and dust resistant though designed to resist moisture, must not be considered completely water-tight. Mounting to walls or racks must be made securely. Avoid locations that are subject to extreme vibration and sway.

Sufficient space must be provided around the analyzer to accommodate the necessary electrical conduit and plumbing connections. The front panel must be allowed to pull out for possible service access to all components of the enclosure. Refer to the system/analyzer outline drawings for dimensions.

Regardless of configuration, the analyzer/system must be installed on a level surface with sufficient space allocated on either side for personnel and test equipment access. Subject to the foregoing, the Analyzer/System should be

Teledyne Analytical Instruments

2-1