Teledyne OFFST/INVRT: <ENT> Standard: <ESC>, Operation, Model 7300A, Imp: CO2 FR:Ø TO:1Ø %

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 64
Image 64
Sel rng to set appl: —> Ø1 Ø2 Ø3 CAL <—

4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

In the autoranging mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument automatically shifts to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument switches to the lower range. A correspond- ing shift in the DC concentration output, and in the range ID outputs, will be noticed.

The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the contaminant concentration detected. If the concen- tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc (20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read beyond the full-scale setting until amplifier saturation is reached. Below amplifier saturation, the over range readings are accurate UNLESS the application uses linearization over the selected range.

To program the ranges, you must first perform the four steps indicated at the beginning of section 4.8 Programming. You will then be in the MAIN MENU and selecting application function screen.

Sel rng to set appl: —> Ø1 Ø2 Ø3 CAL <—

Use the UP/DOWN key to increment/decrement the range number to 01, 02, 03, or CAL, and Enter.

Imp: CO2

FR:Ø TO:1Ø %

Use the UP/DOWN key to increment the respective parameters as de- sired.

Use the ENTER to move from Imp: (impurity) to ppm or % to FR: (from- lower end of ranges), and TO: (to upper-end ranges).

Last Enter will accept the values and the screen will display

OFFST/INVRT: <ENT>

Standard: <ESC>

ESC if your application is standard. If your application requires OFFSET or Reversal output, ENTER will set your output as OFFSET or REVERSAL. (See special function setup section 4.9 for more information).

. (See note below.) Repeat for each range you want to set.

4-28

Teledyne Analytical Instruments

Page 64
Image 64
Teledyne manual OFFST/INVRT: <ENT> Standard: <ESC>, Operation, Model 7300A, Teledyne Analytical Instruments