Teledyne 7300A manual 4.3.10Zero Offset Adjustment, Infrared Gas Analyzer, Operation

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 49
Image 49
4.3.10Zero Offset Adjustment

Infrared Gas Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

6160

7320

8640

91280

102360

The response time listed above can and will change depending on the application and they merely serve to illustrate the effect of the digital filter. The digital filter disengages if there is an upset that is more than 5% of full scale. As the reading comes within the range, the filter becomes active once again.

4.3.10Zero Offset Adjustment

The software in this instrument provides a way to enter an offset on the zero operation of the analyzer. If the background gas of the process is different than the zero calibration gas being used, the reading will have an offset that will be constant throughout its working range. Thus, the need to provide an offset when the instrument is being zeroed.

How to access the offset function:

To access this function, Scroll through the menu and select OFFSET. Press the Enter key to see the following screen:

Enter zero offset:

0.00 %

The offset value can be modified by using the Up/Down keys. Next section shows how to select this value. Suffice to say that whatever value you enter, will be automatically added to the reading. Thus, if you entered -0.10 %, at the end of the zero the display will show -0.10 %.

When the instrument enters the zero mode in AUTO, the instrument will do the work of bringing the reading back to zero plus the offset value that was entered. If you chose MANual zero mode, then you must adjust the zero of the instrument the corresponding section of the manual but with one difference: instead of bringing the display to read zero, you must make the display read zero plus the value entered as offset.

Teledyne Analytical Instruments

4-13

Page 49
Image 49
Teledyne 7300A manual 4.3.10Zero Offset Adjustment, Infrared Gas Analyzer, Operation, Teledyne Analytical Instruments