4 Operation | Model 7300A | |
|
|
|
|
|
|
|
|
|
Note: If all three ranges are currently defined for different applica- tion gases, then the above screen does not display (because mode must be manual). Instead, the VFD goes directly to the following screen.
If above screen displays, use the UP/DOWN arrow keys to Select MANU- AL, and press Enter.
Select range to run
Use the UP/DOWN keys to select the range: 01, 02, 03, or CAL. Then press Enter.
Gas use: CO2
Range: Ø — 10 %
Use the ENTER key to move the range to
Press Escape to return to the previous screen to select or define another range.
Press Enter to return the to the Analyze function.
4.6.2 Auto Screen
Autoranging will automatically set to the application that has at least two range setup with the same gases.
In the autoranging mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument automatically shifts to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument switches to the lower range. A correspond- ing shift in the DC concentration output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the contaminant concentration detected. If the concen- tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the
Teledyne Analytical Instruments |