Teledyne 7300A manual Span Cal, Infrared Gas Analyzer, Operation, 4.4.2.1Auto Mode Spanning

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 55
Image 55
4.4.2 Span Cal

Infrared Gas Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

b.Check for leaks downstream from the sensor, where contamina- tion may be leaking into the system.

c.Check flowmeter to ensure that the flow is no more than 200SCCM

d.Check temperature controller board.

e.Check gas temperature.

If none of the above as indicated, the sensor may need to be replaced. Check warranty, and contact Analytical Instruments Customer Service.

4.4.2 Span Cal

The Span button on the front panel is used to span calibrate the analyzer. Span calibration can be performed in either the automatic or manual mode.

CAUTION: If you are spanning the Cal Range by itself (multiple application analyzers only), use manual mode zeroing.

If you want to calibrate ALL of the ranges at once (multiple application analyzers only), use auto mode spanning in the Cal Range.

Make sure the span gas is flowing to the instrument.

4.4.2.1Auto Mode Spanning

Observe all precautions in sections 4.4 and 4.4.2, above. Scroll SPAN and enter the span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Use the UP/ DOWN key to toggle between AUTO and MAN span settling. Stop when AUTO appears, blinking, on the display.

Select span mode: AUTO

Enter to move to the next screen.

Span Val: 2Ø.ØØ % <ENT> To begin span

Use UP/DOWN key to change the span setting value.

Teledyne Analytical Instruments

4-19

Page 55
Image 55
Teledyne 7300A Span Cal, Infrared Gas Analyzer, Operation, 4.4.2.1Auto Mode Spanning, Teledyne Analytical Instruments