2 Installation | Model 7300A | |
|
|
|
|
|
|
|
|
|
placed as close to the sample point as possible and bolted to its supporting surface. When installed as a system with enclosure
2.2.1 User Connections
All user connections are on the back of the equipment panel and appear in the outline diagram in the back of the manual or addendum to the manual.
2.2.2 Electrical Power Connections
The standard power requires a supply of
Warning: Primary power to the system should not be supplied until all customer wiring is inspected properly by
2.2.3 Calibration Gases
The system may require a supply of clean, oil and particulate free air for use as zero gas.
For accurate calibration, the analyzer requires a blended gas mixture, typically
Do not restrict the bypass, sample or reference vents of the analyzer (and sample system when provided). All lines must vent to a stable safe
Teledyne Analytical Instruments |